An Automated Blood Pressure Display for Self-Measurement in Patients With Chronic Kidney Disease (iHealth Track): Device Validation Study

UDC.coleccionInvestigaciónes_ES
UDC.departamentoCiencias da Saúdees_ES
UDC.grupoInvGrupo de Investigación Saúde e Podoloxía (GISAP)es_ES
UDC.issue4es_ES
UDC.journalTitleJMIR mHealth and uHealthes_ES
UDC.volume8es_ES
dc.contributor.authorMazoteras-Pardo, Victoria
dc.contributor.authorBecerro-de-Bengoa-Vallejo, Ricardo
dc.contributor.authorLosa Iglesias, Marta Elena
dc.contributor.authorLópez-López, Daniel
dc.contributor.authorCalvo-Lobo, César
dc.contributor.authorRodríguez Sanz, David
dc.contributor.authorMartínez-Jiménez, Eva María
dc.contributor.authorPalomo-López, Patricia
dc.date.accessioned2020-04-21T15:13:07Z
dc.date.available2020-04-21T15:13:07Z
dc.date.issued2020-04
dc.description.abstract[Abstract] Background: Hypertension is a global public health issue and is closely related to chronic kidney disorder (CKD). In people with CKD, strict monitoring of blood pressure is an important part of therapy. Objective: The aim of this research was to validate the iHealth Track blood pressure monitoring device for patients with CKD according to the European Society of Hypertension International Protocol 2010 (ESH-IP2). Methods: In total, 33 patients who received hemodialysis in Plasencia participated in the study. There were 9 successive measurements made, which conformed to the ESH-IP2. We calculated the differences between the standard reference device (Omron M3 Intellisense) and the test device (iHealth Track) for blood pressure and heart rate values. For 99 total comparisons of paired measurements, we classified differences into various categories (≤5 mmHg, ≤10 mmHg, and ≤15 mmHg for blood pressure; ≤3, ≤5, and ≤8 beats per minute for heart rate). Results: In 90 of 99 systolic blood pressure and 89 of 99 diastolic blood pressure comparisons between the devices, measurement differences were within 5 mmHg. In 81 of 99 heart rate comparisons between the devices, measurement differences were within 3 beats per minute. The mean differences between the test and reference standard measurements were 3.27 (SD 2.99) mmHg for systolic blood pressure, 3.59 (SD 4.55) mmHg for diastolic blood pressure, and 2.18 (SD 2.75) beats per minute for heart rate. We also observed that for both systolic and diastolic blood pressure, 31 of 33 participants had at least two of three comparisons between the devices with measurement differences less than 5 mmHg. For heart rate, 28 of 33 patients had at least two of three comparisons between the devices with measurement differences less than 3 beats per minute. Conclusions: To our knowledge, this is the first study to show that iHealth Track meets the requirements of the ESH-IP2 in patients with CKD. Therefore, the iHealth Track is suitable for use in renal patients.es_ES
dc.identifier.citationMazoteras-Pardo V, Becerro-De-Bengoa-Vallejo R, Losa-Iglesias ME, López-López D, Calvo-Lobo C, Rodríguez-Sanz D, Martínez-Jiménez EM, Palomo-López P An Automated Blood Pressure Display for Self-Measurement in Patients With Chronic Kidney Disease (iHealth Track): Device Validation Study JMIR Mhealth Uhealth 2020;8(4):e14702 DOI: 10.2196/14702 PMID: 32238337es_ES
dc.identifier.issn2291-5222
dc.identifier.urihttp://hdl.handle.net/2183/25382
dc.language.isoenges_ES
dc.publisherJMIR Publicationses_ES
dc.relation.urihttps://doi.org/10.2196/14702es_ES
dc.rightsAtribución 3.0 Españaes_ES
dc.rights.accessRightsopen accesses_ES
dc.rights.urihttp://creativecommons.org/licenses/by/3.0/es/*
dc.subjectiHealth Trackes_ES
dc.subjectValidationes_ES
dc.subjectBlood pressurees_ES
dc.subjectHeart ratees_ES
dc.subjectInternational protocoles_ES
dc.titleAn Automated Blood Pressure Display for Self-Measurement in Patients With Chronic Kidney Disease (iHealth Track): Device Validation Studyes_ES
dc.typejournal articlees_ES
dspace.entity.typePublication
relation.isAuthorOfPublication80e827fb-79b5-45af-9756-581db1fa7b73
relation.isAuthorOfPublication.latestForDiscovery80e827fb-79b5-45af-9756-581db1fa7b73

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
MazoterasPardo_Victoria_2020_An_Automated_Blood_Pressure_Display.pdf
Size:
638.33 KB
Format:
Adobe Portable Document Format
Description: