Radon Mitigation Approach in a Laboratory Measurement Room

Loading...
Thumbnail Image

Identifiers

Publication date

Authors

Blanco Rodriguez, Patricia
Cos Juez, Francisco Javier de

Advisors

Other responsabilities

Journal Title

Bibliographic citation

Blanco-Rodríguez, P.; Fernández-Serantes, L.A.; Otero-Pazos, A.; Calvo-Rolle, J.L.; de Cos Juez, F.J. Radon Mitigation Approach in a Laboratory Measurement Room. Sensors 2017, 17,5; 1090.

Type of academic work

Academic degree

Abstract

[Abstract] Radon gas is the second leading cause of lung cancer, causing thousands of deaths annually. It can be a problem for people or animals in houses, workplaces, schools or any building. Therefore, its mitigation has become essential to avoid health problems and to prevent radon from interfering in radioactive measurements. This study describes the implementation of radon mitigation systems at a radioactivity laboratory in order to reduce interferences in the different works carried out. A large set of radon concentration samples is obtained from measurements at the laboratory. While several mitigation methods were taken into account, the final applied solution is explained in detail, obtaining thus very good results by reducing the radon concentration by 76%.

Description

Rights

Reconocimiento 4.0 Internacional
Reconocimiento 4.0 Internacional

Except where otherwise noted, this item's license is described as Reconocimiento 4.0 Internacional