QardioArm Upper Arm Blood Pressure Monitor Against Omron M3 Upper Arm Blood Pressure Monitor in Patients With Chronic Kidney Disease : A Validation Study According to the European Society of Hypertension International Protocol Revision 2010

Use this link to cite
http://hdl.handle.net/2183/24562Collections
- Investigación (FEP) [494]
Metadata
Show full item recordTitle
QardioArm Upper Arm Blood Pressure Monitor Against Omron M3 Upper Arm Blood Pressure Monitor in Patients With Chronic Kidney Disease : A Validation Study According to the European Society of Hypertension International Protocol Revision 2010Author(s)
Date
2019-12Citation
Mazoteras-Pardo V, Becerro-De-Bengoa-Vallejo R, Losa-Iglesias ME, López-López D, Rodríguez-Sanz D, Casado-Hernández I, Calvo-Lobo C, Palomo-López P QardioArm Upper Arm Blood Pressure Monitor Against Omron M3 Upper Arm Blood Pressure Monitor in Patients With Chronic Kidney Disease: A Validation Study According to the European Society of Hypertension International Protocol Revision 2010 J Med Internet Res 2019;21(12):e14686 DOI: 10.2196/14686 PMID: 31789600
Abstract
[Abstract] Background: Hypertension is considered as a main risk factor for chronic kidney disease development and progression. Thus,
the control and evaluation of this disease with new software and devices are especially important in patients who suffer from
chronic kidney disease.
Objective: This study aimed to validate the QardioArm mobile device, which is used for blood pressure (BP) self-measurement
in patients who suffer from chronic kidney disease, by following the European Society of Hypertension International Protocol 2
(ESH-IP2) guidelines.
Methods: A validation study was carried out by following the ESH-IP2 guidelines. A sample of 33 patients with chronic kidney
disease self-measured their BP by using the QardioArm and Omron M3 Intellisense devices. Heart rate (HR), diastolic BP, and
systolic BP were measured.
Results: The QardioArm fulfilled the ESH-IP2 validation criteria in patients who suffered from chronic kidney disease.
Conclusions: Thus, this study is considered as the first validation using a wireless upper arm oscillometric device connected
to an app to measure BP and HR meeting the ESH-IP2 requirements in patients who suffer from chronic kidney disease. New
validation studies following the ESH-IP2 guidelines should be carried out using different BP devices in patients with specific
diseases.
Keywords
Blood pressure
Hypertension
Kidney disease
Mobile apps
Software validation
Hypertension
Kidney disease
Mobile apps
Software validation
Editor version
Rights
Atribución 3.0 España
ISSN
1438-8871