Validation of a clinical-grade assay to measure donor-derived cell-free DNA in solid organ transplant recipients

Ver/ abrir
Use este enlace para citar
http://hdl.handle.net/2183/18466
A non ser que se indique outra cousa, a licenza do ítem descríbese como Atribución-NoComercial-SinDerivadas 3.0 España
Coleccións
- Investigación (FCS) [1284]
Metadatos
Mostrar o rexistro completo do ítemTítulo
Validation of a clinical-grade assay to measure donor-derived cell-free DNA in solid organ transplant recipientsAutor(es)
Data
2016-10-07Cita bibliográfica
Grskovic M, Hiller DJ, Eubank LA, et al. Validation of a clinical-grade assay to measure donor-derived cell-free DNA in solid organ transplant recipients. J Mol Diag. 2016;18(6):890-902
Resumo
[Abstract] The use of circulating cell-free DNA (cfDNA) as a biomarker in transplant recipients offers advantages over invasive tissue biopsy as a quantitative measure for detection of transplant rejection and immunosuppression optimization. However, the fraction of donor-derived cfDNA (dd-cfDNA) in transplant recipient plasma is low and challenging to quantify. Previously reported methods to measure dd-cfDNA require donor and recipient genotyping, which is impractical in clinical settings and adds cost. We developed a targeted next-generation sequencing assay that uses 266 single-nucleotide polymorphisms to accurately quantify dd-cfDNA in transplant recipients without separate genotyping. Analytical performance of the assay was characterized and validated using 1117 samples comprising the National Institute for Standards and Technology Genome in a Bottle human reference genome, independently validated reference materials, and clinical samples. The assay quantifies the fraction of dd-cfDNA in both unrelated and related donor-recipient pairs. The dd-cfDNA assay can reliably measure dd-cfDNA (limit of blank, 0.10%; limit of detection, 0.16%; limit of quantification, 0.20%) across the linear quantifiable range (0.2% to 16%) with across-run CVs of 6.8%. Precision was also evaluated for independently processed clinical sample replicates and is similar to across-run precision. Application of the assay to clinical samples from heart transplant recipients demonstrated increased levels of dd-cfDNA in patients with biopsy-confirmed rejection and decreased levels of dd-cfDNA after successful rejection treatment. This noninvasive clinical-grade sequencing assay can be completed within 3 days, providing the practical turnaround time preferred for transplanted organ surveillance.
Versión do editor
Dereitos
Atribución-NoComercial-SinDerivadas 3.0 España
ISSN
1525-1578
1943-7811
1943-7811